Therac-25Therac-25 was a radiation therapy machine produced by Atomic Energy of Canada Limited. It was involved with at least six known accidents between 1985 and 1987, in which patients were given massive overdoses. These accidents highlighted the dangers of software control of safety-critical systems (see computer bug).
Problem descriptionThe machine had two treatment modes: When operating in direct electron-beam therapy mode, a low-powered electron beam was emitted directly from the machine. When operating in soft X-ray mode, a beam flattener (in the turntable of the machine) was to be rotated into the path of the beam and caused to emit (safe amount of) X-rays through the use of a much higher electron-beam current (about 100 times of beam power in electron-beam therapy mode). The accidents occurred when the high-energy electron-beam was activated without the beam flattener having been rotated into place. The very high energy electron-beam directly struck the patients causing the feeling of an intense electric shock and the occurrence of thermal and radiation burns. In some cases, the injured patients died later from radiation poisoning. Root causesResearchers who investigated the accidents found several contributing causes. These included the following institutional causes:
The researchers also found several engineering issues:
These incidents have become a standard case study in the history of computing and medicine. See alsoExternal links